Abstract
Background: Insulin-like growth factor 1 (IGF1) is a biomarker with various applications in medicine and also in doping control.
Methods: A liquid chromatography-tandem mass spectrometry (LC-MS/MS) method was developed that employs 15N-IGF1 as an internal standard. The method features urea-based IGF1/IGFBP-complex dissociation which is directly followed by tryptic digestion. Following solidphase extraction (SPE) sample clean-up of the digest, IGF1 is detected by means of two signature peptides that enable quantification of total IGF1 as well as discrimination between IGF1 proteoforms with 'native' and modified or extended N-terminal sequences.
Results: Our method is capable of measuring plasma IGF1 concentrations over the clinically relevant range of 10-1000 ng/mL and was validated according to regulatory guidelines. Comparison with the IDS-iSYS IGF1 immunoassay revealed good correlation (R-2 > 0.97) and no proportional bias between both assays was observed after normalizing the results against the WHO reference standard for IGF1 (02/254). Evaluation of several commercially available IGF1 preparations showed varying responses which were due to inconsistencies in purity and absolute amount of IGF1 present in these products.
Conclusions: Our LC-MS/MS method introduces urea-based dissociation of IGF1/IGFBP-complexes to enable reliable quantification of IGF1 in plasma. Furthermore, the method is able to detect clinically relevant IGF1 levels without an enrichment procedure at the protein-level and thereby minimizes the risk of losing IGF1 proteoforms during sample preparation.
| Original language | English |
|---|---|
| Pages (from-to) | 1905-1912 |
| Number of pages | 8 |
| Journal | Clinical chemistry and laboratory medicine |
| Volume | 56 |
| Issue number | 11 |
| Early online date | 2018 |
| DOIs | |
| Publication status | Published - Nov-2018 |
Keywords
- insulin-like growth factor 1 (IGF1)
- liquid chromatography coupled with tandem mass spectrometry (LC-MS/MS)
- quantification
- CHROMATOGRAPHY/TANDEM MASS-SPECTROMETRY
- FACTOR-I
- GH DEFICIENCY
- IGF-I
- BIOLOGICAL SAMPLES
- HORMONE ABUSE
- QUANTIFICATION
- PROTEINS
- SERUM
- PERFORMANCE