An assessment of computer-assisted personalized sedation: a sedation delivery system to administer propofol for gastrointestinal endoscopy

Daniel J. Pambianco*, Christopher J. Whitten, Annelies Moerman, Michel M. Struys, James F. Martin

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

79 Citations (Scopus)

Abstract

Background: Demand for colonoscopy and EGD procedures is increasing. Impediments to performing these examinations persist, Patients perceive these procedures as unpleasant and painful. The use of suboptimal sedatives results in inefficiency in endoscopy practices. Improving sedation methods utilizing precise control of preferred sedatives may increase patient satisfaction ancl practice efficiency

Objective: Our purpose was to demonstrate the feasibility of computer-assisted personalized sedation (CAPS) for facilitating the precise administration of propofol by endoscopist/nurse teams, achieving minimal to moderate sedation in subjects undergoing routine endoscopies.

Design: Open label, single-center studies.

Setting: Endoscopy clinics in Charlottesville, Virginia, and Gent, Belgium.

Subjects: Twenty-four adults per center; 12 colonoscopies, 12 EGDs.

Interventions: Propofol sedation with CAPS by endoscopist/registered nurse care teams.

Main Outcome Measurements: Sedation level measured by modified observer's assessment of alertness/sedation (MOAA/S), recovery time measured from endoscope removal until Adrete >= 12, dosage of propofol, oxygen saturation, and safety assessments.

Results: Subjects responded to mild tactile and verbal stimuli MOAA/S = 5, 4, 3, or 2) 99% of the time. Mean propofol closes in the United States and Belgium were 65.4 ancl 72.1 mg, respectively mean recovery times were 29 and 10 seconds, respectively. Oxygen desaturation occurred in only 6% of subjects. No device-related adverse events occurred.

Limitation: Open-label design.

Conclusions: Using CAPS, the endoscopist/nurse teams precisely controlled the administration of propofol achieving minimal to moderate sedation in subjects undergoing colonoscopy and EGD procedures. Mean propofol dosage was low and post-procedure recovery times were rapid. The device performed well when operated by the endoscopist/nurse team, with no device-related adverse events.

Original languageEnglish
Pages (from-to)542-547
Number of pages6
JournalGastrointestinal endoscopy
Volume68
Issue number3
DOIs
Publication statusPublished - Sept-2008
Externally publishedYes

Keywords

  • THERAPEUTIC UPPER ENDOSCOPY
  • SCREENING COLONOSCOPY
  • RESPIRATORY ACTIVITY
  • GRAPHIC ASSESSMENT
  • MIDAZOLAM

Fingerprint

Dive into the research topics of 'An assessment of computer-assisted personalized sedation: a sedation delivery system to administer propofol for gastrointestinal endoscopy'. Together they form a unique fingerprint.

Cite this