TY - JOUR
T1 - Assessment of pedicle size in patients with scoliosis using EOS 2D imaging
T2 - a validity and reliability study
AU - Peeters, C. M. M.
AU - van Houten, L.
AU - Kempen, D. H. R.
AU - Wapstra, F. H.
AU - Jutte, P. C.
AU - van den Akker-Scheek, I.
AU - Faber, C.
PY - 2021/4/25
Y1 - 2021/4/25
N2 - Purpose Free-hand pedicle screw insertion methods are widely used for screw insertion during scoliosis surgery. Preoperative knowledge about the pedicle size helps to maximize screw containment and minimize the risk of pedicle breach. Radiographs taken by a biplanar low-dose X-ray device (EOS) have no divergence in the vertical plane. The criterion validity and reliability of preoperative EOS images for pedicle size measurements in patients with idiopathic scoliosis (IS) was investigated in this study. Methods Sixteen patients who underwent surgical treatment for IS were prospectively included. Intra- and extracortical pedicle height and width measurements on EOS images were compared with reconstructed intra-operative 3D images of the isthmus of included pedicles. Secondly, intra- and interobserver reliability of pedicle size measurements on EOS images was determined. Results The total number of analyzed pedicles was 203. The correlation between the EOS and 3D scan measurements was very strong for the intra- and extracortical pedicle height and strong for the intra- and extracortical pedicle width. There are, however, significant, but likely clinically irrelevant differences (mean absolute differences < 0.43 mm) between the two measure methods for all four measurements except for extracortical pedicle height. For pedicles classified as Nash-Moe 0, no significant differences in intra- and extracortical pedicle width were observed. Both intra- and interobserver reliability was excellent for all pedicle size measurements on EOS images. Conclusion The results of this study indicate a good validity and reliability for pedicle size measurements on EOS radiographs. Therefore, EOS radiographs may be used for a preoperative estimation of pedicle size and subsequent screw diameter in patients with IS.
AB - Purpose Free-hand pedicle screw insertion methods are widely used for screw insertion during scoliosis surgery. Preoperative knowledge about the pedicle size helps to maximize screw containment and minimize the risk of pedicle breach. Radiographs taken by a biplanar low-dose X-ray device (EOS) have no divergence in the vertical plane. The criterion validity and reliability of preoperative EOS images for pedicle size measurements in patients with idiopathic scoliosis (IS) was investigated in this study. Methods Sixteen patients who underwent surgical treatment for IS were prospectively included. Intra- and extracortical pedicle height and width measurements on EOS images were compared with reconstructed intra-operative 3D images of the isthmus of included pedicles. Secondly, intra- and interobserver reliability of pedicle size measurements on EOS images was determined. Results The total number of analyzed pedicles was 203. The correlation between the EOS and 3D scan measurements was very strong for the intra- and extracortical pedicle height and strong for the intra- and extracortical pedicle width. There are, however, significant, but likely clinically irrelevant differences (mean absolute differences < 0.43 mm) between the two measure methods for all four measurements except for extracortical pedicle height. For pedicles classified as Nash-Moe 0, no significant differences in intra- and extracortical pedicle width were observed. Both intra- and interobserver reliability was excellent for all pedicle size measurements on EOS images. Conclusion The results of this study indicate a good validity and reliability for pedicle size measurements on EOS radiographs. Therefore, EOS radiographs may be used for a preoperative estimation of pedicle size and subsequent screw diameter in patients with IS.
KW - Pedicle size measurements
KW - EOS
KW - Intra-operative 3D images
KW - Scoliosis
KW - Validity and reliability
U2 - 10.1007/s00586-021-06839-8
DO - 10.1007/s00586-021-06839-8
M3 - Article
C2 - 33895877
SP - 1
EP - 9
JO - European Spine Journal
JF - European Spine Journal
SN - 0940-6719
ER -