Abstract
Background: Current pulsatile pumps for cardiopulmonary bypass (CPB) are far from satisfactory because of the poor pulsatility. This study was undertaken to examine the efficiency of a novel pulsatile catheter pump on pulsatility and its effect on abdominal organ perfusion during CPB. Methods: Twelve pigs weighing 89+/-11 kg were randomly divided into a pulsatile group (n=6) and a non-pulsatile group (n=6). All animals had a CPB for 120 min, aorta clamped for 60 min, temperature down to 32degreesC, and a perfusion flow of 60 mI/kg/min. In the pulsatile group, a 21 Fr intra-aortic pulsatile catheter, which was connected to a 40 mL membrane pump, was placed in the descending aorta and activated by a balloon pump driver during the first 90 minutes of CPB until aortic declamping. Hemodynamics, organ blood flow, body metabolism, and blood trauma were studied during experiments.
Results: Compared with the non-pulsatile group during CPB, the pulsatile group had a higher systolic blood pressure (p
Conclusion: Applying the pulsatile catheter pump in the descending aorta is effective in supplying the pulsatile flow to the abdominal organs and results in improved abdominal organ perfusion during the ischemic phase of CPB.
| Original language | English |
|---|---|
| Pages (from-to) | 35-43 |
| Number of pages | 9 |
| Journal | International journal of artificial organs |
| Volume | 28 |
| Issue number | 1 |
| Publication status | Published - Jan-2005 |
Keywords
- pulsatile pump
- abdominal organs
- cardiopulmonary bypass
- ENERGY EQUIVALENT PRESSURE
- NONPULSATILE PERFUSION
- BLOOD-FLOW
- IN-VIVO
- CONTROVERSY
Fingerprint
Dive into the research topics of 'Augmentation of abdominal organ perfusion during cardiopulmonary bypass with a novel intra-aortic pulsatile catheter pump'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver