Automated analysis of small animal PET studies through deformable registration to an atlas

  • Daniel F. Gutierrez
  • , Habib Zaidi*
  • *Corresponding author for this work

    Research output: Contribution to journalArticleAcademicpeer-review

    19 Citations (Scopus)

    Abstract

    This work aims to develop a methodology for automated atlas-guided analysis of small animal positron emission tomography (PET) data through deformable registration to an anatomical mouse model.

    A non-rigid registration technique is used to put into correspondence relevant anatomical regions of rodent CT images from combined PET/CT studies to corresponding CT images of the Digimouse anatomical mouse model. The latter provides a pre-segmented atlas consisting of 21 anatomical regions suitable for automated quantitative analysis. Image registration is performed using a package based on the Insight Toolkit allowing the implementation of various image registration algorithms. The optimal parameters obtained for deformable registration were applied to simulated and experimental mouse PET/CT studies. The accuracy of the image registration procedure was assessed by segmenting mouse CT images into seven regions: brain, lungs, heart, kidneys, bladder, skeleton and the rest of the body. This was accomplished prior to image registration using a semi-automated algorithm. Each mouse segmentation was transformed using the parameters obtained during CT to CT image registration. The resulting segmentation was compared with the original Digimouse atlas to quantify image registration accuracy using established metrics such as the Dice coefficient and Hausdorff distance. PET images were then transformed using the same technique and automated quantitative analysis of tracer uptake performed.

    The Dice coefficient and Hausdorff distance show fair to excellent agreement and a mean registration mismatch distance of about 6 mm. The results demonstrate good quantification accuracy in most of the regions, especially the brain, but not in the bladder, as expected. Normalized mean activity estimates were preserved between the reference and automated quantification techniques with relative errors below 10 % in most of the organs considered.

    The proposed automated quantification technique is reliable, robust and suitable for fast quantification of preclinical PET data in large serial studies.

    Original languageEnglish
    Pages (from-to)1807-1820
    Number of pages14
    JournalEuropean Journal of Nuclear Medicine and Molecular Imaging
    Volume39
    Issue number11
    DOIs
    Publication statusPublished - Nov-2012

    Keywords

    • PET/CT
    • Small animals
    • Quantification
    • Deformable registration
    • Atlas
    • IMAGE REGISTRATION
    • RAT-BRAIN
    • 3-DIMENSIONAL REGISTRATION
    • EMISSION-TOMOGRAPHY
    • DOSE CALCULATIONS
    • MRI REGISTRATION
    • MOUSE PHANTOM
    • CT
    • SCANNER
    • MICE

    Fingerprint

    Dive into the research topics of 'Automated analysis of small animal PET studies through deformable registration to an atlas'. Together they form a unique fingerprint.

    Cite this