Does the patch fit the stoma? A study on peristoma geometry and patch use in laryngectomized patients

  • E. B. van der Houwen*
  • , T. A. van Kalkeren
  • , W. J. Post
  • , F. J. M. Hilgers
  • , B. F. A. M. van der Laan
  • , G. J. Verkerke
  • *Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

16 Citations (Scopus)

Abstract

Objective: To determine the geometry of (peri) stomas of laryngectomised patients in relation to patch use. This data will enable improvement of tracheostoma interfaces, specifically addressing patients currently unable to use stoma patches. The low use of Heat-and-Moisture-Exchange (HME) filters and (hands-free) speech valves, although very important to the quality of life of laryngectomised patients, is mainly attributed to poor fit of the adhesive patches to the stoma site. Current patch shapes are not based upon an objective (peri) stoma geometry because this geometry is unknown.

Design: Observational anthropometric study of the (peri) stoma of laryngectomised patients.

Setting: Ten hospitals or institutes in eight countries.

Participants: About 191 laryngectomised patients, at least 1 year post operative.

Main outcome measures: (Peri) stomas were photographed and measured. Patients completed a questionnaire on patch-use. Concavity of commercially available patches was measured.

Results: In countries with a financial reimbursement system 58% of the patients use patches, compares to only 9% in other countries. Patches stay in situ for an average of 33.3 h. Patch and non-patch users differ on five out of ten measured geometrical parameters. Most striking differences are that patch users have much shallower peristomas (13 versus 18 mm), and stomas far more parallel to the anterior neck plane. The deepest commercially available patch is only 7 mm deep.

Conclusions: This study provides detailed (peri) stoma geometry data of a divers population, and for the first time in relation to patch-use. It reveals a serious mismatch between patients and patches. With these data new patches can be developed that could dramatically improve rehabilitation after laryngectomy.

Original languageEnglish
Pages (from-to)235-241
Number of pages7
JournalClinical Otolaryngology
Volume36
Issue number3
DOIs
Publication statusPublished - 2011
EventBME 2011 Conference - , Netherlands
Duration: 20-Jan-201121-Jan-2011