Electronic portal images (EPIs) based position verification for the breast simultaneous integrated boost (SIB) technique

Research output: Contribution to journalArticleAcademicpeer-review

8 Citations (Scopus)

Abstract

Background and purpose: To develop a method based on electronic portal images (EPIs) for the position verification of breast cancer patients that are treated with a simultaneous integrated boost (SIB) technique.

Method: 3D setup errors of the breast outline and the thoracic wall were determined from EPIs of the tangential treatment fields and anterior posterior (AP) verification field. The method was verified with repeated CT scans of 38 patients with an average setup error larger than 5 mm.

Result: The 3D position deviation of the boost volume can best be determined from the position deviation of the breast outline in the ventrodorsal direction and the thoracic wall in the lateral and longitudinal directions from the tangential and AP EPIs. The method gives an average overestimation of the deviation of the boost volume in the ventrodorsal, lateral and longitudinal directions by 28%, 20% and 6%, respectively and an average underestimation of the deviation of the whole breast by 32%, 17% and 39%.

Conclusions: The described method is superior to using tangential EPIs only and is recommended for position verification of breast cancer patients that are treated with a SIB technique if no Cone beam CT (CBCT) or fiducial markers can be used. (C) 2011 Elsevier Ireland Ltd. All rights reserved. Radiotherapy and Oncology 102 (2012) 108-114

Original languageEnglish
Pages (from-to)108-114
Number of pages7
JournalRadiotherapy and Oncology
Volume102
Issue number1
DOIs
Publication statusPublished - Jan-2012

Keywords

  • Breast conserving therapy
  • Electronic portal imaging
  • EPID
  • Position verification
  • Radiotherapy