Experimental assessment of a new direct fixation implant for artificial limbs

P. K. Tomaszewski, B. Lasnier, G. Hannink, G. J. Verkerke*, N. Verdonschot

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

13 Citations (Scopus)

Abstract

An alternative to prosthetic socket rehabilitation of patients with transfemoral amputation is realized by means of direct skeletal fixation devices, offering significant improvements in mobility and comfort. However, strain shielding due to high stiffness of these metal-based implants causes considerable and progressive bone loss. To overcome this phenomenon a new concept of a direct fixation implant, in form of a collared metallic pin articulating inside a polymer intramedullary part, was developed. In this study we used experimental and finite element techniques to assess whether the novel concept produced a more physiological strain distribution in the bone as compared to a standard titanium implant.

Cortical strains were measured experimentally on seven human cadaver femora, both intact and implanted with a generic standard implant and the new implant. Three load configurations were considered, simulating: heel strike, toe off and one leg stance. A finite element model derived from computed tomography data was used to calculate strains in intact bone and bone with generic models of the two implant types.

Significant strain shielding occurred around both implant types, albeit that for the novel design strain shielding was generally less (p

In conclusion, the study showed that the new implant has a potential to increase distal load transfer to the femur and reduce strain shielding as compared with the standard implant. Collar-cortex contact is an important aspect and requires further attention when developing the surgical technique. The encouraging results obtained in this study justify further development of this concept in order to improve the quality and applicability of direct skeletal fixation devices for patients requiring a transfemoral amputation. (C) 2013 Elsevier Ltd. All rights reserved.

Original languageEnglish
Pages (from-to)77-85
Number of pages9
JournalJournal of the Mechanical Behavior of Biomedical Materials
Volume21
DOIs
Publication statusPublished - May-2013

Keywords

  • Strain shielding
  • Finite element analysis
  • Osseointegrated percutaneous implant
  • Trans-femoral amputation
  • Direct bone attachment
  • FINITE-ELEMENT MODELS
  • STRAIN DISTRIBUTION
  • FEMORAL STEMS
  • IN-VITRO
  • BONE FAILURE
  • PROSTHESES
  • FEMUR
  • RISK
  • OSSEOINTEGRATION
  • REHABILITATION