NE102A Plastic Scintillator Response to He-3 Ions

  • Fatemeh Jafari
  • , Ali Akbar Mehmandoost-Khajeh-Dad*
  • , Maysam Mohammadi-Dadkan
  • , Nima Ghal-Eh
  • , Hamideh Yazdandoust
  • *Corresponding author for this work

    Research output: Contribution to journalArticleAcademicpeer-review

    194 Downloads (Pure)

    Abstract

    Plastic scintillators are widely used in detecting nuclear radiation due to their low construction cost, the ability to be produced in nearly any shape and size and relatively fast time response, among which NE102A (or its equivalents, BC400 and EJ-212) is one of the most famous and widely used scintillators in the focal plane of the particle magnetic spectrometers. In this study, the response of a large NE102A scintillator to He-3 ions was investigated in the energy range of 55-87 MeV. The experimental data were collected from the measurements carried out at the accelerator center of University of Groningen, the Netherlands (KVI-CART). The results of this research, acceptably consistent with previous published experimental data, confirmed that the response of the NE102A scintillator to He-3 ions with energies more than 10 MeV is approximately linear.

    Original languageEnglish
    Pages (from-to)761-767
    Number of pages7
    JournalPhysics of Particles and Nuclei Letters
    Volume16
    Issue number6
    DOIs
    Publication statusPublished - Nov-2019

    Keywords

    • He-3
    • NE102A
    • BC400
    • EJ-212
    • plastic scintillator
    • non-linear response
    • optical photon transport
    • DETECTORS

    Fingerprint

    Dive into the research topics of 'NE102A Plastic Scintillator Response to He-3 Ions'. Together they form a unique fingerprint.

    Cite this