Pharmacy compounded pilocarpine: An adequate solution to overcome shortage of pilogel® discs for sweat testing in patients with cystic fibrosis

P Mian*, J M Maurer, D J Touw, M J Vos, B L Rottier

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

1 Citation (Scopus)
81 Downloads (Pure)

Abstract

To respond to shortage of pilocarpine discs due to CE-licensing problems a pharmacy compounded pilocarpine HCL solution was developed and validated for use in CF diagnosis. The aim of this study was to compare the results from a pharmacy compounded pilocarpine HCL solution versus Pilogel® discs for the measurements of sweat chloride concentrations. Ten pediatric and adult patients with CF underwent a sweat test using both Pilogel® discs and pilocarpine HCL solution. The average difference between both methods was -3.25 mmol/L (95% Limits of Agreement: -7.19 [95% CI: -9.19;-5.19] and 0.69 [95% CI: -1.31;2.69] mmol/L. Passing-Bablok regression showed that zero was enclosed with the 95% CI of the calculated intercept (0.15 [95% CI: -1.70;1.42] mmol/L). These data show a good agreement in chloride concentrations obtained using the two pilocarpine products. Therefore, the pharmacy compounded pilocarpine HCL solution can be used as an alternative for Pilogel® discs during times of shortage.

Original languageEnglish
Pages (from-to)126-131
Number of pages6
JournalJournal of Cystic Fibrosis
Volume23
Issue number1
Early online date27-Sept-2023
DOIs
Publication statusPublished - Jan-2024

Fingerprint

Dive into the research topics of 'Pharmacy compounded pilocarpine: An adequate solution to overcome shortage of pilogel® discs for sweat testing in patients with cystic fibrosis'. Together they form a unique fingerprint.

Cite this