Phosphorylcholine coating of extracorporeal circuits provides natural protection against blood activation by the material surface

F De Somer*, K Francois, W van Oeveren, J Poelaert, D De Wolf, T Ebels, G Van Nooten

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

98 Citations (Scopus)

Abstract

Objective: The aim of this study is to evaluate the use of a new coating, mimicking the outer cell membrane, in paediatric cardiac surgery.

Methods: Two groups of ten patients with a body weight below 8 kg, undergoing elective cardiac operations for different congenital anomalies. were prospectively enrolled in this study. in one group the whole extracorporeal circuit, including the cannulas, was coated with phosphorylcholine (PC). Ln the second group the same circuit was used without coating. Platelet activation (thromboxane B2 (TXB2), beta -thromboglobulin (beta TG)), activation of the coagulation system (F1 + 2), leukocyte activation (CD11b/CD18) and terminal complement activation (TCC) were analyzed pre-cardiopulmonary bypass (CPB), at 15, 60 min of CPB, at the end of CPB, 20 min post CPB and at postoperative day 1 and 6. Results. No statistical differences were found for F1 + 2 and CD11b/CD 18. After onset of CPB mean levels of TCC remained stable in the PC roup whereas an increase was observed in the control group. During CPB beta TG values in both groups increased to a maximum at the end of CPB. Within groups the increase in beta TG levels during CPB was statistically significant (P <0.05) from baseline in the control group starting from 60 min of CPB whereas no statistical difference was observed in the PC group. After the start of CPB TXB2 mean levels increased to 405 +/- 249 pg/ml in the PC group vs. 535 +/- 224 pg/ml in the control group. After this initial increase there was a small decline in the PC group with further increase. This was in contrast to the control group were TXB2 levels further increased up to a mean of 718 +/- 333 pg/ml at the end of CPB (P = 0.016). Conclusions: Phosphorylcholine coating had a favourable effect on blood platelets, which is most obvious after studying the changes during cardiopulmonary bypass. A steady increase of TXB2 and TG was observed in the control group, whereas plateau formation was observed in the phosphorylcholine group. Clinically, this effect may contribute to reduced blood loss and less thromboembolic complications. Complement activation is lower in the coated,group. (C) 2000 Elsevier Science B.V.All rights reserved.

Original languageEnglish
Pages (from-to)602-606
Number of pages5
JournalEUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY
Volume18
Issue number5
Publication statusPublished - Nov-2000

Keywords

  • phosphorylcholine coating
  • paediatric surgery
  • cardiopulmonary bypass
  • platelets
  • complement
  • CARDIOPULMONARY BYPASS
  • PLATELET-ADHESION
  • FIBRINOGEN
  • MEMBRANES
  • POLYMERS

Fingerprint

Dive into the research topics of 'Phosphorylcholine coating of extracorporeal circuits provides natural protection against blood activation by the material surface'. Together they form a unique fingerprint.

Cite this