Abstract
Most image analysis systems (IAS) use black-and-white cameras. However, true color IASs are considered to be useful for quantification of immunohistologically stained structures. Using a true color IAS, we evaluated two methods of segmentation for quantification of area percentage of staining: one using fixed, preset thresholds and one using thresholds interactively set per image. Furthermore, the effect of shading correction pens evaluated, and measurements in both color and black-and-white mode were compared. The results of segmentation with fixed thresholds did not differ significantly from those of control percentages, established by interactive morphometry using a grid, which served as reference. Interactive segmentation was significantly different from the reference (t test, P =.0002). The effect of shading correction teas negligible. Measurements with and without this procedure correlated highly (r =.99, P
| Original language | English |
|---|---|
| Pages (from-to) | 357-364 |
| Number of pages | 8 |
| Journal | Analytical and Quantitative Cytology and Histology |
| Volume | 16 |
| Issue number | 5 |
| Publication status | Published - Oct-1994 |
Keywords
- IMAGE PROCESSING, COMPUTER-ASSISTED
- IMMUNOHISTOCHEMISTRY
- BREAST-CANCER
- EPITHELIAL VOLUME
- FLOW-CYTOMETRY
- OVARIAN-TUMORS
- RECEPTORS
- ESTROGEN
- QUANTITATION
- EXPRESSION
Fingerprint
Dive into the research topics of 'Quantification of area percentage of immunohistochemical staining by true color image analysis with application of fixed thresholds'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver