Abstract
Background: The short and long term variability of the interrupter technique was assessed to determine whether interrupter resistance is a stable individual characteristic over time. The effect of field and standardised measurement conditions on the within-subject variability of the interrupter technique was also examined.
Methods: The interrupter technique was studied under field and standardised conditions in children aged 3-6 years. Under field conditions, five investigators performed the measurements using two different measurement devices in random sequence. Both short term (20-30 minutes) and long term variability (median 38 days) were assessed in 32 children. Under standardised conditions, a single investigator conducted all measurements using a single device; the repeated measurements were conducted at the same time of day in a familiar quiet classroom. Long term variability (median 11 days) was estimated in 15 children. Within-subject standard deviations were estimated by analysis of variance with adjustment for the effects of different investigators and measurement devices on within-subject variability under field conditions.
Results: Under field conditions within-subject standard deviations for short and long term variability were 0.10 kPa/l/s (adjusted 0.10 kPa/l/s) and 0.13 kPa/l/s (adjusted 0.14 kPa/l/s), respectively. Under standardised conditions the within-subject standard deviation for long term variability was 0.10 kPa/l/s.
Conclusions: Measurement of interrupter resistance under field conditions only slightly increased the within-subject variability compared with standardised conditions. The results indicate that interrupter resistance is a stable individual characteristic over a period of some weeks.
Original language | English |
---|---|
Pages (from-to) | 761-764 |
Number of pages | 4 |
Journal | Thorax |
Volume | 58 |
Issue number | 9 |
Publication status | Published - 1-Sept-2003 |
Keywords
- LUNG-FUNCTION
- PRESCHOOL-CHILDREN
- AIRWAY-RESISTANCE
- REFERENCE VALUES
- YOUNG-CHILDREN
- BRONCHODILATOR
- CHALLENGE
- DEVICE