Washing of stored red blood cells by an autotransfusion device before transfusion

R. de Vroege*, W. R. Wildevuur, J. A. G. Muradin, D. Graves, W. van Oeveren

*Corresponding author for this work

    Research output: Contribution to journalArticleAcademicpeer-review

    29 Citations (Scopus)

    Abstract

    Background and Objectives The use of an autotransfusion device to wash blood of the incision site is increasing. After washing, this blood is retransfused without side effects caused by activated plasma factors and cell release products. This procedure could be extended to washing of donor blood, which may be particularly useful for red blood cells (RBCs) stored for more than 4 weeks that contain high concentrations of free haemoglobin, potassium, lactate and other metabolites. It is not known whether stored RBCs can withstand the cell washing procedure with the use of an autotransfusion device, while keeping their primary functions intact. The objective of this study was to determine the quality of RBCs, after cell washing in comparison to untreated RBCs.

    Materials and Methods RBCs were studied in terms of integrity (free haemoglobin), stored energy (2,3-diphosphoglycerate, adenosine triphosphate), metabolites (lactate, potassium) and physical characteristics (osmotic resistance, aggregability, deformability).

    Results After washing, free lactate and potassium were significantly reduced as compared to the levels before washing. The osmotic resistance of RBCs slightly improved, whereas aggregation capacity reduced after washing. Fifteen per cent of haemoglobin was lost during washing. The deformability and free haemoglobin levels remained unchanged.

    Conclusion Washing stored blood before transfusion may be of benefit, because the waste products are effectively removed from the stored RBC.

    Original languageEnglish
    Pages (from-to)130-135
    Number of pages6
    JournalVox Sanguinis
    Volume92
    Issue number2
    DOIs
    Publication statusPublished - 2007

    Keywords

    • aggregation
    • autotransfusion device
    • deformability
    • RBCs
    • transfusion
    • washing
    • CARDIAC-SURGERY
    • STORAGE
    • QUALITY
    • AGE

    Cite this