A novel liquid chromatography-tandem mass spectrometry method is described for the quantitative determination of the endogenous CYP 3A4/5 marker 4 beta-hydroxycholesterol in human K(2)-EDTA plasma. It is based on alkaline hydrolysis to convert esterified to free 4 beta-hydroxycholesterol, followed by analyte extraction from plasma by hexane and purification of the hexane extract by normal-phase solid-phase extraction. The analyte is chromatographically separated from endogenous isobaric plasma oxysterols and excess cholesterol by a 16-min reversed-phase gradient on a C18 column; detection is performed by atmospheric pressure photoionization tandem mass spectrometry in the positive ion mode, using toluene as a dopant.
Using 400 mu l of plasma, 4 beta-hydroxycholesterol can be quantified in the concentration range 10.0-250 nM. Validation results show that the method is sufficiently selective towards endogenous plasma sterols and capable of quantifying the analyte with good precision and accuracy. The analyte is sufficiently stable in all relevant matrices and solvents; the addition of the anti-oxidant butylated hydroxytoluene to prevent in vitro formation of 4 beta-hydroxycholesterol from cholesterol during storage or analysis is not necessary, provided that long-term frozen storage of plasma occurs at -70 degrees C. (C) 2011 Elsevier B.V. All rights reserved.