Considerations in the use of different spirometers in epidemiological studies

Edith B Milanzi, Gerard H Koppelman, Marieke Oldenwening, Sonja Augustijn, Bernadette Aalders-de Ruijter, Martijn Farenhorst, Judith M Vonk, Marjan Tewis, Bert Brunekreef, Ulrike Gehring*

*Corresponding author voor dit werk

Onderzoeksoutput: ArticleAcademicpeer review

18 Citaten (Scopus)
364 Downloads (Pure)

Samenvatting

BACKGROUND: Spirometric lung function measurements have been proven to be excellent objective markers of respiratory morbidity. The use of different types of spirometers in epidemiological and clinical studies may present systematically different results affecting interpretation and implication of results. We aimed to explore considerations in the use of different spirometers in epidemiological studies by comparing forced expiratory volume in 1 s (FEV1) and forced vital capacity (FVC) measurements between the Masterscreen pneumotachograph and EasyOne spirometers. We also provide a correction equation for correcting systematic differences using regression calibration.

METHODS: Forty-nine volunteers had lung function measured on two different spirometers in random order with at least three attempts on each spirometer. Data were analysed using correlation plots, Bland and Altman plots and formal paired t-tests. We used regression calibration to provide a correction equation.

RESULTS: The mean (SD) FEV1 and FVC was 3.78 (0.63) L and 4.78 (0.63) L for the Masterscreen pneumotachograph and 3.54 (0.60) L and 4.41 (0.83) L for the EasyOne spirometer. The mean FEV1 difference of 0.24 L and mean FVC difference of 0.37 L between the spirometers (corresponding to 6.3 and 8.4% difference, respectively) were statistically significant and consistent between younger (< 30 years) and older volunteers (> 30 years) and between males and females. Regression calibration indicated that an increase of 1 L in the EasyOne measurements corresponded to an average increase of 1.032 L in FEV1 and 1.005 L in FVC in the Masterscreen measurements.

CONCLUSION: Use of different types of spirometers may result in significant systematic differences in lung function values. Epidemiological researchers need to be aware of these potential systematic differences and correct for them in analyses using methods such as regression calibration.

Originele taal-2English
Artikelnummer39
Aantal pagina's8
TijdschriftEnvironmental health
Volume18
Nummer van het tijdschrift1
DOI's
StatusPublished - 25-apr.-2019