TY - JOUR
T1 - Early sepsis recognition
T2 - a pilot study using a rapid high-multiplex host-response mRNA diagnostic test
AU - Lu, Jingyi
AU - ter Voert, Michèle A.
AU - Ünal, Mehtap
AU - Whitfield, Natalie N.
AU - Liesenfeld, Oliver
AU - ter Maaten, Jan C.
AU - Sweeney, Timothy E.
AU - Bouma, Hjalmar R.
N1 - Publisher Copyright:
© The Author(s) 2025.
PY - 2025/2/21
Y1 - 2025/2/21
N2 - Background: Early sepsis diagnosis is essential to allow timely initiation of adequate care. The TriVerity™ Test, performed on the Myrna™ Instrument, is the first rapid high-multiplex host-response mRNA diagnostic test that supports clinical decision-making by evaluating the likelihood of bacterial and/or viral infections and severity of illness. We present findings of the first, proof of concept, real-world evaluation in an emergency department (ED). Methods: Blood was collected in PAXgene® Blood RNA tubes from adult patients visiting the ED with suspicion of infection between 4th December 2023 and 22nd January 2024. TriVerity was performed within 1 h (RNA extraction and amplification of 29 host mRNAs using LAMP technology on the Myrna Instrument within approximately 30 min). TriVerity generates three diagnostic scores (likelihood of bacterial infection, viral infection, and illness severity), each classified into five discrete interpretation bands (very low, low, moderate, high, and very high). Post hoc chart reviews were performed after hospital discharge to clinically adjudicate the infection status (consensus and forced adjudication). Results: Among 60 patients, there were 20 (33%) bacterial infections, 15 (25%) viral infections, 11 (18%) bacterial–viral coinfections and 14 (23%) did not have an infection under forced adjudication. Under consensus adjudication, bacterial results demonstrated 95% rule-in specificity and 95% rule-out sensitivity. Viral results demonstrated rule-in specificity 100% and 92% rule-out sensitivity. Since only three patients were admitted to the ICU or died in this cohort, we cannot draw firm conclusions about the predictive value of the test for severe endpoints. Conclusions: TriVerity is a rapid whole-blood host-response test to reliably detect the presence or absence of bacterial and/or viral infections, as well as to assess illness severity in patients presenting to the ED. Its quick turnaround time aligns with the ED workflow, offering timely insights for clinical decision-making. Results from upcoming large-scale validation studies will provide more detailed results on the diagnostic and prognostic accuracy of the test.
AB - Background: Early sepsis diagnosis is essential to allow timely initiation of adequate care. The TriVerity™ Test, performed on the Myrna™ Instrument, is the first rapid high-multiplex host-response mRNA diagnostic test that supports clinical decision-making by evaluating the likelihood of bacterial and/or viral infections and severity of illness. We present findings of the first, proof of concept, real-world evaluation in an emergency department (ED). Methods: Blood was collected in PAXgene® Blood RNA tubes from adult patients visiting the ED with suspicion of infection between 4th December 2023 and 22nd January 2024. TriVerity was performed within 1 h (RNA extraction and amplification of 29 host mRNAs using LAMP technology on the Myrna Instrument within approximately 30 min). TriVerity generates three diagnostic scores (likelihood of bacterial infection, viral infection, and illness severity), each classified into five discrete interpretation bands (very low, low, moderate, high, and very high). Post hoc chart reviews were performed after hospital discharge to clinically adjudicate the infection status (consensus and forced adjudication). Results: Among 60 patients, there were 20 (33%) bacterial infections, 15 (25%) viral infections, 11 (18%) bacterial–viral coinfections and 14 (23%) did not have an infection under forced adjudication. Under consensus adjudication, bacterial results demonstrated 95% rule-in specificity and 95% rule-out sensitivity. Viral results demonstrated rule-in specificity 100% and 92% rule-out sensitivity. Since only three patients were admitted to the ICU or died in this cohort, we cannot draw firm conclusions about the predictive value of the test for severe endpoints. Conclusions: TriVerity is a rapid whole-blood host-response test to reliably detect the presence or absence of bacterial and/or viral infections, as well as to assess illness severity in patients presenting to the ED. Its quick turnaround time aligns with the ED workflow, offering timely insights for clinical decision-making. Results from upcoming large-scale validation studies will provide more detailed results on the diagnostic and prognostic accuracy of the test.
KW - Acute infection
KW - Acutelines
KW - Diagnosis
KW - Host response
KW - mRNA
KW - Prognosis
KW - Sepsis
KW - TriVerity
UR - http://www.scopus.com/inward/record.url?scp=85218503328&partnerID=8YFLogxK
U2 - 10.1186/s40635-025-00735-x
DO - 10.1186/s40635-025-00735-x
M3 - Article
AN - SCOPUS:85218503328
SN - 2197-425X
VL - 13
JO - Intensive Care Medicine Experimental
JF - Intensive Care Medicine Experimental
M1 - 21
ER -