Samenvatting
Objectives: We sought to investigate factors that affect the difference between the peak inspiratory pressure measured at the Y-piece under dynamic flow conditions and plateau pressure measured under zero-flow conditions (resistive pressure) during pressure controlled ventilation across a range of endotracheal tube sizes, respiratory mechanics, and ventilator settings. Design: In vitro study. Setting: Research laboratory. Patients: None. Interventions: An in vitro bench model of the intubated respiratory system during pressure controlled ventilation was used to obtain the difference between peak inspiratory pressure measured at the Y-piece under dynamic flow conditions and plateau pressure measured under zero-flow conditions across a range of endotracheal tubes sizes (3.0–8.0 mm). Measurements were taken at combinations of pressure above positive end-expiratory pressure (10, 15, and 20 cm H2O), airway resistance (no, low, high), respiratory system compliance (ranging from normal to extremely severe), and inspiratory time at constant positive end-expiratory pressure (5 cm H2O). Multiple regression analysis was used to construct models predicting resistive pressure stratified by endotracheal tube size. Measurements and Main Results: On univariate regression analysis, respiratory system compliance (β –1.5; 95% CI, –1.7 to –1.4; p < 0.001), respiratory system resistance (β 1.7; 95% CI, 1.5–2.0; p < 0.001), pressure above positive end-expiratory pressure (β 1.7; 95% CI, 1.4–2.0; p < 0.001), and inspiratory time (β –0.7; 95% CI, –1.0 to –0.4; p < 0.001) were associated with resistive pressure. Multiple linear regression analysis showed the independent association between increasing respiratory system compliance, increasing airway resistance, increasing pressure above positive end-expiratory pressure, and decreasing inspiratory time and resistive pressure across all endotracheal tube sizes. Inspiratory time was the strongest variable associated with a proportional increase in resistive pressure. The contribution of airway resistance became more prominent with increasing endotracheal tube size. Conclusions: Peak inspiratory pressures measured during pressure controlled ventilation overestimated plateau pressure irrespective of endotracheal tube size, especially with decreased inspiratory time or increased airway resistance.
| Originele taal-2 | English |
|---|---|
| Pagina's (van-tot) | E47-E51 |
| Aantal pagina's | 5 |
| Tijdschrift | Pediatric critical care medicine : a journal of the Society of Critical Care Medicine and the World Federation of Pediatric Intensive and Critical Care Societies |
| Volume | 21 |
| Nummer van het tijdschrift | 1 |
| Vroegere onlinedatum | 1-nov.-2019 |
| DOI's | |
| Status | Published - jan.-2020 |
Vingerafdruk
Duik in de onderzoeksthema's van 'Effect of Endotracheal Tube Size, Respiratory System Mechanics, and Ventilator Settings on Driving Pressure'. Samen vormen ze een unieke vingerafdruk.Citeer dit
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS