Factors that influence standard automated perimetry test results in glaucoma: Test reliability, technician experience, time of day, and season

Francisco G. Junoy Montolio, Christiaan Wesselink, Marijke Gordijn, Nomdo M. Jansonius*

*Corresponding author voor dit werk

OnderzoeksoutputAcademicpeer review

94 Citaten (Scopus)
664 Downloads (Pure)

Samenvatting

PURPOSE. To determine the influence of several factors on standard automated perimetry test results in glaucoma.

METHODS. Longitudinal Humphrey field analyzer 30-2 Swedish interactive threshold algorithm data from 160 eyes of 160 glaucoma patients were used. The influence of technician experience, time of day, and season on the mean deviation (MD) was determined by performing linear regression analysis of MD against time on a series of visual fields and subsequently performing a multiple linear regression analysis with the MD residuals as dependent variable and the factors mentioned above as independent variables. Analyses were performed with and without adjustment for the test reliability (fixation losses and false-positive and false-negative answers) and with and without stratification according to disease stage (baseline MD).

RESULTS. Mean follow-up was 9.4 years, with on average 10.8 tests per patient. Technician experience, time of day, and season were associated with the MD. Approximately 0.2 dB lower MD values were found for inexperienced technicians (P <0.001), tests performed after lunch (P <0.001), and tests performed in the summer or autumn (P <0.001). The effects of time of day and season appeared to depend on disease stage. Independent of these effects, the percentage of false-positive answers strongly influenced the MD with a 1 dB increase in MD per 10% increase in false-positive answers.

CONCLUSIONS. Technician experience, time of day, season, and the percentage of false-positive answers have a significant influence on the MD of standard automated perimetry. (Invest Ophthalmol Vis Sci. 2012;53:7010-7017) DOI:10.1167/iovs.12-10268

Originele taal-2English
Pagina's (van-tot)7010-7017
Aantal pagina's8
TijdschriftInvestigative ophthalmology & visual science
Volume53
Nummer van het tijdschrift11
DOI's
StatusPublished - okt.-2012