Samenvatting
Objectives: Regulators of circadian rhythm, including melatonin, influence fundamental biological processes. Measuring the melatonin metabolite 6-sulfatoxymelatonin in urine can estimate melatonin production. 6-sulfatoxymelatonin is mainly analyzed by immunoassays, but these methods are hampered by cross-reactivity and poor reproducibility when used to analyze small molecules. Therefore, we validated a high-throughput liquid chromatography with tandem mass spectrometry (LC-MS/MS) method to quantify 6-sulfatoxymelatonin in urine. We evaluated age-dependent 24-h excretion of 6-sulfatoxymelatonin into urine and the biological variation of urinary excretion in healthy individuals.
Methods: The online solid phase extraction method combined with LC-MS/MS was validated according to international guidelines, and used to measure the excretion of 6-sulfatoxymelatonin into urine of 240 healthy individuals. Biological variation of 6-sulfatoxymelatonin excretion was examined in 10 healthy individuals.
Results: Urinary 6-sulfatoxymelatonin results were well within the validation criteria (interassay coefficient of variation:
Conclusions: This MS-based method enables straightforward, reproducible, and sensitive quantification of 6-sulfatoxymelatonin in urine. Urinary 6-sulfatoxymelatonin levels decreased with age. Biological variation of 6-sulfatoxymelatonin excretion into urine was high between subjects and lower within subjects, indicating that repeated measurements of 6-sulfatoxymelatonin in 24-h urine are needed in future studies.
Originele taal-2 | English |
---|---|
Pagina's (van-tot) | 187-195 |
Aantal pagina's | 9 |
Tijdschrift | Clinical chemistry and laboratory medicine |
Volume | 59 |
Nummer van het tijdschrift | 1 |
Vroegere onlinedatum | 16-jul.-2020 |
DOI's | |
Status | Published - 1-jan.-2021 |