Numerical simulation of the pulsating catheter pump: A left ventricular assist device

GJ Verkerke*, D Mihaylov, AA Geertsema, J Lubbers, G Rakhorst

*Bijbehorende auteur voor dit werk

OnderzoeksoutputAcademicpeer review

12 Citaten (Scopus)

Samenvatting

The pulsating catheter (PUCA) pump, a left ventricular assist device, consists of a hydraulically or pneumatically driven membrane pump, extracorporeally placed and mounted to a valved catheter. The catheter is introduced into an easily accessible artery and positioned with its distal tip in the left ventricle. Blood is aspirated from the left ventricle during systole and ejected into the ascending aorta during diastole. A numerical model of the PUCA pump has been developed to determine the internal diameter of the PUCA pump catheter that allows a certain blood flow. The model considers a limitation of mechanical blood damage and determines the accompanying pressure and flow profile for driving the pump. For a flow of 5 L/min, a catheter with an internal diameter of at least 6.95 mm is required. For 3 L/min, the minimal diameter is 5.50 mm. The latter catheter can be introduced in the axillary artery, the former via the aorta during an open thorax surgical procedure. To validate the numerical model, 2 different PUCA pump configurations were tested in vitro. Results showed a good resemblance between model and in vitro behavior of the PUCA pump.

Originele taal-2English
Pagina's (van-tot)924-931
Aantal pagina's8
TijdschriftARTIFICIAL ORGANS
Volume23
Nummer van het tijdschrift10
StatusPublished - okt-1999