Stability, survival, and tolerability of a 4.5-mm-wide bone-anchored hearing implant: 6-month data from a randomized controlled clinical trial

Rik C Nelissen*, Christine A den Besten, Emmanuel A M Mylanus, Myrthe K S Hol

*Corresponding author voor dit werk

Onderzoeksoutput: ArticleAcademicpeer review

40 Citaten (Scopus)

Samenvatting

The objective of this study was to compare the stability, survival, and tolerability of 2 percutaneous osseointegrated titanium implants for bone conduction hearing: a 4.5-mm diameter implant (test) and a 3.75-mm diameter implant (control). Fifty-seven adult patients were included in this randomized controlled clinical trial. Sixty implants were allocated in a 2:1 (test-control) ratio. Follow-up visits were scheduled at 7, 14, 21, and 28 days; 6 and 12 weeks; and 6 months. At every visit, implant stability quotient (ISQ) values were recorded by means of resonance frequency analysis (RFA) and skin reactions were evaluated according to the Holgers classification. Implants were loaded with the bone conduction device at 3 weeks. Hearing-related quality of life was evaluated using the Abbreviated Profile of Hearing Aid Benefit (APHAB), the Glasgow Benefit Inventory (GBI), and the Glasgow Health Status Inventory (GHSI). ISQ values were statistically significantly higher for the test implant compared to the control implant. No implants were lost and soft tissue reactions were comparable for both implants. Positive results were reported in the hearing-related quality of life questionnaires. These 6-month results indicate that both implants and their corresponding hearing devices are safe options for hearing rehabilitation in patients with the appropriate indications. Loading at 3 weeks did not affect the stability of either implant.

Originele taal-2English
Pagina's (van-tot)105-111
Aantal pagina's7
TijdschriftEuropean Archives of Oto-Rhino-Laryngology
Volume273
DOI's
StatusPublished - jan.-2016
Extern gepubliceerdJa

Vingerafdruk

Duik in de onderzoeksthema's van 'Stability, survival, and tolerability of a 4.5-mm-wide bone-anchored hearing implant: 6-month data from a randomized controlled clinical trial'. Samen vormen ze een unieke vingerafdruk.

Citeer dit