Samenvatting
We compared eleven OCT devices in their ability to quantify retinal layer thicknesses under different signal-strength conditions, using a commercially available phantom eye. We analyzed a medium-intensity 50 µm layer in an identical manner for all devices, using the provided log-scale images and a reconstructed linear-scale tissue reflectivity metric. Thickness measurements were highly comparable when the data were analyzed in an identical manner. With optimal signal strength, the thickness of the 50 µm layer was overestimated by a mean of 4.3 µm in the log-scale images and of 2.7 µm in the linear-scale images.
| Originele taal-2 | English |
|---|---|
| Pagina's (van-tot) | 1306-1315 |
| Aantal pagina's | 10 |
| Tijdschrift | Biomedical optics express |
| Volume | 11 |
| Nummer van het tijdschrift | 3 |
| DOI's | |
| Status | Published - 1-mrt.-2020 |