TY - JOUR
T1 - The Aptima HPV assay fulfills the cross-sectional clinical and reproducibility criteria of international guidelines for human papillomavirus test requirements for cervical screening
AU - Heideman, D A M
AU - Hesselink, A T
AU - van Kemenade, F J
AU - Iftner, T
AU - Berkhof, J
AU - Topal, F
AU - Agard, D
AU - Meijer, C J L M
AU - Snijders, P J F
PY - 2013/11
Y1 - 2013/11
N2 - The Aptima HPV assay (Hologic Gen-Probe, San Diego, CA) is an FDA-approved assay for detecting human papillomavirus (HPV) E6/E7 mRNA from 14 high-risk HPV types. This study evaluated the clinical performance of the Aptima HPV assay for cervical intraepithelial neoplasia of grade 2 or worse (CIN2+), relative to the high-risk HPV GP5+/GP6+ PCR, in a cross-sectional clinical equivalence analysis using the noninferiority score test with cervical samples from population-based screening, i.e., 69 cervical scraping samples from women with CIN2+ and 843 from women without evidence of CIN2+. In addition, intralaboratory reproducibility over time and interlaboratory agreement of the Aptima HPV assay results were assessed with another set of 548 cervical samples. The Aptima HPV assay showed a clinical sensitivity for CIN2+ of 94.2% (95% confidence interval [CI], 85.5 to 97.8%) and a clinical specificity for CIN2+ of 94.5% (95% CI, 92.8 to 95.9%); by comparison, these figures were 97.1% (95% CI, 89.1 to 99.3%) (67/69 samples) and 93.6% (95% CI, 91.7 to 95.0%) (785/839 samples), respectively, for GP5+/GP6+ PCR. The clinical sensitivity and specificity of the Aptima HPV assay were noninferior to those of GP5+/GP6+ PCR (P = 0.039 and 0.00016, respectively). In addition, high reproducibility of the Aptima HPV assay, as reflected by the intralaboratory reproducibility over time of 96.0% (95% CI, 94.4 to 97.3%) (526/548 samples; kappa = 0.89) and interlaboratory agreement of 96.7% (95% CI, 95.4 to 98.1%) (531/548 samples; kappa = 0.91), was found. Altogether, these data show that the Aptima HPV assay meets the cross-sectional clinical and reproducibility criteria of the international guidelines for HPV test requirements for cervical screening. Longitudinal data are needed to ensure that the long-term negative predictive value of this mRNA assay is similar to those of validated HPV DNA tests.
AB - The Aptima HPV assay (Hologic Gen-Probe, San Diego, CA) is an FDA-approved assay for detecting human papillomavirus (HPV) E6/E7 mRNA from 14 high-risk HPV types. This study evaluated the clinical performance of the Aptima HPV assay for cervical intraepithelial neoplasia of grade 2 or worse (CIN2+), relative to the high-risk HPV GP5+/GP6+ PCR, in a cross-sectional clinical equivalence analysis using the noninferiority score test with cervical samples from population-based screening, i.e., 69 cervical scraping samples from women with CIN2+ and 843 from women without evidence of CIN2+. In addition, intralaboratory reproducibility over time and interlaboratory agreement of the Aptima HPV assay results were assessed with another set of 548 cervical samples. The Aptima HPV assay showed a clinical sensitivity for CIN2+ of 94.2% (95% confidence interval [CI], 85.5 to 97.8%) and a clinical specificity for CIN2+ of 94.5% (95% CI, 92.8 to 95.9%); by comparison, these figures were 97.1% (95% CI, 89.1 to 99.3%) (67/69 samples) and 93.6% (95% CI, 91.7 to 95.0%) (785/839 samples), respectively, for GP5+/GP6+ PCR. The clinical sensitivity and specificity of the Aptima HPV assay were noninferior to those of GP5+/GP6+ PCR (P = 0.039 and 0.00016, respectively). In addition, high reproducibility of the Aptima HPV assay, as reflected by the intralaboratory reproducibility over time of 96.0% (95% CI, 94.4 to 97.3%) (526/548 samples; kappa = 0.89) and interlaboratory agreement of 96.7% (95% CI, 95.4 to 98.1%) (531/548 samples; kappa = 0.91), was found. Altogether, these data show that the Aptima HPV assay meets the cross-sectional clinical and reproducibility criteria of the international guidelines for HPV test requirements for cervical screening. Longitudinal data are needed to ensure that the long-term negative predictive value of this mRNA assay is similar to those of validated HPV DNA tests.
KW - Adult
KW - Early Detection of Cancer/methods
KW - Female
KW - Humans
KW - Middle Aged
KW - Molecular Diagnostic Techniques/methods
KW - Papillomaviridae/classification
KW - Papillomavirus Infections/diagnosis
KW - Reproducibility of Results
KW - Sensitivity and Specificity
KW - Uterine Cervical Dysplasia/diagnosis
U2 - 10.1128/JCM.01517-13
DO - 10.1128/JCM.01517-13
M3 - Article
C2 - 23985912
SN - 0095-1137
VL - 51
SP - 3653
EP - 3657
JO - Journal of Clinical Microbiology
JF - Journal of Clinical Microbiology
IS - 11
ER -