The clinical relevance of dry powder inhaler performance for drug delivery

Pascal Demoly*, Paul Hagedoorn, Anne H. de Boer, Henderik W. Frijlink

*Corresponding author voor dit werk

OnderzoeksoutputAcademicpeer review

123 Citaten (Scopus)
298 Downloads (Pure)

Samenvatting

Background: Although understanding of the scientific basis of aerosol therapy with dry powder inhalers (DPIs) has increased, some misconceptions still persist. These include the beliefs that high resistance inhalers are unsuitable for some patients, that extra fine (

Objectives: This article aims to clarify the complex inter-relationships between inhaler design and resistance, inspiratory flow rate (IFR), FPF, lung deposition and clinical outcomes, as a better understanding may result in a better choice of DPI for individual patients.

Methods: The various factors that determine the delivery of drug particles into the lungs are reviewed. These include aerodynamic particle size distribution, the inspiratory manoeuvre, airway geometry and the three basic principles that determine the site and extent of deposition: inertial impaction, sedimentation and diffusion. DPIs are classed as either dependent or independent of inspiratory flow rate and vary in their internal resistance to inspiration. The effects of these characteristics on drug deposition in the airways are described using data from studies directly comparing currently available inhaler devices.

Results: Clinical experience shows that most patients can use a high resistance DPI effectively, even during exacerbations. Particles in the aerodynamic size range from 1.5-5 mu m are shown to be optimal, as particles 5 mu m may impact on the oropharynx. For DPIs with a constant FPF at all flow rates, less of the delivered dose reaches the central and peripheral lung when the flow rate increases, risking underdosing of the required medication. In contrast, flow rate-dependent inhalers increase their FPF output at higher flow rates, which compensates for the greater impaction on the upper airways as flow rate increases.

Conclusions: The technical characteristics of different inhalers and the delivery and deposition of the fine particle dose to the lungs may be important additional considerations to help the physician to select the most appropriate device for the individual patient to optimise their treatment. (C) 2014 Published by Elsevier Ltd.

Originele taal-2English
Pagina's (van-tot)1195-1203
Aantal pagina's9
TijdschriftRespiratory Medicine
Volume108
Nummer van het tijdschrift8
DOI's
StatusPublished - aug.-2014