Samenvatting
The application of Bond-Elut Certify solid-phase extraction columns to the systematic toxicological analysis of whole blood was evaluated. The reproducibility of the extraction was tested with thirteen drugs varying in physico-chemical properties. Analysis was performed with capillary gas chromatography with nitrogen-selective detection. The recoveries were reproducible, as long as other limiting factors, e.g., chromatographic behaviour or volatility, do not play a significant role. The effect of limiting chromatographic behaviour was studied in more detail with the more sensitive mass spectrometry with selected-ion monitoring after converting the extracted morphine into its ditrimethylsilyl derivative.
| Originele taal-2 | English |
|---|---|
| Pagina's (van-tot) | 87-95 |
| Aantal pagina's | 9 |
| Tijdschrift | Journal of Chromatography A |
| Volume | 674 |
| Nummer van het tijdschrift | 1-2 |
| Status | Published - 15-jul.-1994 |